compressed Air Volume is Reduced
- Piston cups in pump are worn. Use the Rebuild Kit to replace.
(P/N95-0069)
- Leak in Regulator. Replace Regulator Assembly (95-0178).
- If the compressor is running constantly, the unloader solenoid may be sticking open. Replace solenoid (95-0177 see
section for complete list of part numbers available).
- Water bottles not sealing allowing air to escape. Replace gaskets (P/N 95-0156) or bottles (P/N 95-0140QT).
- Gasket on handpiece is worn or not tight. Tighten handpiece adapter firmly or replace gasket.
- Air leaks in tubing. Troubleshoot to locate leak and replace tubing and/or fittings.
compressor will not Start
- If the compressor has a pungent electrical smell, the compressor may be stalling when attempting to start. Problem is
usually check valve allowing air back in and back pressuring the heads. Remove check valve clean out and clean or
replace Check Valve (P/N 95-0161).
- Start up capacitor may need replacement (P/N 95-0063 see section for full list of available part numbers).
- Motor overload protection switch may have tripped. Let motor cool off for two hours and the compressor will
automatically restart.
- Low voltage power input. Verify 120 Volts or 230 Volts with multimeter. Correct lower voltage situation or switch to a
dedicated outlet. For 120 Volts use optional power conditioner.
Troubleshooting the Purge Pump
- Low flow caused by kinked tubing. Open back panel and adjust tubing to remove any kinks or tight bends. Verify that
slides on pump head are opened to maximum position.
- Low flow caused by amalgam separator being full. Empty bowl to remove sediment.
- No flow caused by pump not running our bound. Open up pump and adjust tubing; run pump to clear bind.
- Leak inside caused by tear or cut in hose. Replace tubing.