As an alternate, the bottles may be filled with a solution of 10 parts water and 1 part mouthwash that contains alcohol.
For more information on this protocol, please see abstract entitled “A Chemical Treatment Regimen to Reduce Bacterial
Contamination in Dental Water Lines” in the May 1997 issue of JADA. If using another type of disinfectant, follow the
instructions provided with your disinfectant for effective control of the growth of biofilms in the water line.
3.2.2 Syringe Tips
For reusable tips follow the manufacturer’s suggested guidelines for tip cleaning.
3.2.3 Sterilization
The Triton™ Advanced Dental System® and the Attachments offered with it (Air/Water Syringe, Saliva Ejector, and High
Volume Evacuator) are not provided sterile. The air/water syringe and suction handpieces are designed for use with
disposable tips. If using reusable tips supplied by others, they must be properly sterilized before each use according to their
manufacturer’s instructions. In addition to the disinfection of these parts, a disposable barrier sleeve must be used to cover
the air/water syringe, saliva ejector and high volume evacuator (HVE). The disposable barrier sleeve must be discarded and
replaced between each patient.
Follow all manufacturer’s instructions for sterilization of any optional accessories and dental handpieces made by others.
3.2.4 Silestone™ care & Maintenance
Follow the guidelines listed below to keep your Silestone™ composite solid surface worktop looking as good as new.
Routine care
Most dirt and stains
Use soapy water or ammonia-based cleaner.
Water marks
Wipe with damp cloth; towel dry.
Disinfecting
Occasionally wipe surfaces with diluted household
bleach (1 part water/1 part bleach).
3.2.5 Solid Trap
WARninG – infectious Materials
The suction system may contain infectious materials. Always wear
personal protective equipment and follow appropriate protocol.
Check the solid trap filter and if partially obstructed or full replace
with new filter. To replace filter, remove lid, lift out filter and dispose of
according to hazardous waste protocol.