ADVAnceD DenTAl SySTeMS
®
2025M | 2025M/DV | 2025M/AR
99
65-2025M Rev. B
Category
Time Needed/
Minutes
Frequency
Daily Weekly Monthly Annually
3 yrs
5 yrs or
greater
Air/Water Syringe
Install syringe rebuild kit pN# 95-0012 or syringe button
repair kit pN# 95-0089, then lubricate button o-rings pN#
95-0085 (grease)
30
X
Replace tubing pN# 95-0214=50”, pN# 95-0213=60”
30
X
Fiber Optics
Verify bulb is operational, Replace if burn out
10
X
Check integrity of tubing and examine small wires for
potential breakage
10
X
Miscellaneous Electrical
Inspect input wiring cord and internal wiring harnesses and
ensure no damage or fraying
15
X
Optional Suction Instruments and Accessories
Clean and lubricate suction valve lever o-ring and
disconnect o-ring. p/N 95-0085 (grease)
10
X
Inspect, clean and lubricate solids trap lid o-ring
10
X
Inspect all suction tubing for leaks and replace if necessary
10
X
Replace suction lever o-rings and connector barb o-rings
p/Ns 95-0040 and 95-4034
15
X
Replace solids trap lid o-ring pN 95-0091L
15
X
Cart Systems
Clean wheels and casters of all debris with warm water and
disinfectant, then lubricate.
30
X