Verify Mechanical integrity and Safe Use
After unpacking the unit, look for any damage that could affect safe and proper operation. Ensure there are no missing parts
or accessories. Carefully inspect components for tightness and the absence of cracks, excessive wear or frayed electrical
cords. If there is any adverse damage or the lack of mechanical integrity that would prevent the safe use or operation of the
system, it should not be used until repair or replacements correcting the defects are made.
1.5 Dual Voltage conversion
iMPORTAnT nOTe:
This system is capable of operation from either a
120volt/60 Hz or a 230volt/50 Hz power source.
CAUTION: All selector switches must be properly selected for the correct
voltage and frequency input for proper operation and to prevent damage to
electrical components.
nOTe:
The cart system has three selector switches and an optional fourth switch
if equipped with a KaVo electric motor. Two voltage selectors are located on top
plate, and the other two where the power cord enters the cart at the back of the
cart.
Step 1:
Ensure cart is unplugged from the power source. Worktop is held in place
with commercial grade Velcro. Carefully remove top – note that Velcro is very
strong and will require a firm grasp.
Step 2:
Locate the voltage switch located at the rear of the top panel and select
either “230V” for 230V/50 Hz or “115V” for 120V/60 Hz. Use a small flat
screwdriver to select the desired voltage
Step 3:
If the cart is equipped with a KaVo brand Electric Motor, it will be
necessary to switch it to the correct voltage as well. Locate the Voltage Selector
Switch Located at the right side of the top panel. Use a small flat screwdriver to
select the desired voltage.
Step 4.
Locate the power cord inlet plate on the back of the cart. Remove the plate
by removing the two screws. Flip both of the voltage selectors on the back of the
cart to the desired voltage. Replace the power inlet plate.
Step 5.
Replace the power cord plug with the correct type if necessary. The cart has been supplied with both a 115 Volt
NEMA type U.S. plug and Schuko 230 Volt European type Plug.
Step 2
Step 3
Step 4