Care should be taken to use only manufacturer-recommended cables, power supplies and electrical accessories with
the Triton™ Advanced Dental System®.
If the user or a third-party supplier offers cables, external power supplies and electrical accessories for use with the
Triton™ Advanced Dental System® and they are not listed or referenced in this manual, it is the responsibility of that
third-party supplier to determine compliance with the standards and tests in the following tables.
• The use of electrical cables and accessories other than those specified in this manual or referenced documents
may result in increased electromagnetic emissions from the Triton™ Advanced Dental System® or decreased
electromagnetic immunity of the Triton™ Advanced Dental System®.
Guidance and Manufacturer’s Declaration – electromagnetic emissions
The Triton™ Advanced Dental System® is intended for use in the electromagnetic environment specified below. The customer or the end user of the Triton™
Advanced Dental System® should assure that it is used in such an environment.
emissions Test
compliance
electromagnetic environment - Guidance
RF Emissions - CISPR 11
(Radiated & Conducted)
Group 1
The Triton™ Advanced Dental System® uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions - CISPR 11
(Radiated & Conducted)
Class B
The Triton™ Advanced Dental System® is suitable for use in all establishments,
including domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic purposes.
Harmonic Emissions
EN/IEC 61000-3-2
Class A
Voltage fluctuations/
Flicker Emissions
EN/IEC 61000-3-3
Complies