Guidance and Manufacturer’s Declaration – electromagnetic immunity
The Triton™ Advanced Dental System® is intended for use in the electromagnetic environment specified below. The customer or the end user of the Triton™
Advanced Dental System® unit should assure it is used only in such an environment.
immunity Test
en/iec 60601
Test level
compliance level
intended electromagnetic environment
Electromagnetic Discharge (ESD)
EN/IEC 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient/burst
EN/IEC 61000-4-4
± 2kV for power supply lines
± 1kV for input/output lines
± 2kV for power supply
lines
± 1kV for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
EN/IEC 61000-4-5
± 1kV differential mode (line-
line)
± 2kV common mode (line-
earth)
± 1kV differential mode
(line-line)
± 2kV common mode
(line-earth)
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short interruptions and
voltage variations on power supply
input lines
EN/IEC 61000-4-11
<5% UT (>95% dip in UT) for
0.5 cycle
40% UT (60% dip in UT) for
5 cycles
70% UT (30% dip in UT) for
25 cycles
<5% UT (>95% dip in UT) for
5 seconds
<5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 seconds
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
Triton™ Advanced Dental System® requires continued
operation during power mains interruptions, it is
recommended that the Triton™ Advanced Dental
System® be powered from an uninterruptible power
supply or a battery.
Power frequency (50/60Hz)
magnetic field
EN/IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Note UT is the a.c. mains voltage prior to application of the test level.