3.0 Maintenance
3.1 equipment care & inspection
The equipment should be thoroughly cleaned and disinfected after use each day to maintain optimal performance. All
components should be maintained in proper working order and replaced as required.
3.1.1 inspection
The following visual checks and inspections should be performed before operating the unit:
• Look for mechanical damage that could affect safe operation
• Ensure there are no missing parts or accessories.
• Inspect components for tightness, cracks, wear or frayed electrical cords.
• Should mechanical or other damage be evident that would affect the safe use or operation of the system, it should not
be used until repair or replacements correcting the defects are made.
3.1.2 Preventative Maintenance
See Schedule 10.1 for a recommended preventative maintenance schedule for the delivery system.
3.2 cleaning, Disinfection & Sterilization
3.2.1 Disinfecting the Dual closed clean Water System & Handpiece Tubing
It is recommended to use the DentaPure Cartridge in the clean water system bottles to disinfect the water
system. The DP365 is designed for use for one year without requiring daily maintenance. This cartridge is
manufactured by MLRB International and is available for purchase directly from ASI. To order online, visit
asimedical.net/store or call 800.566.9953 and request part number 95-0281.
To use, the cartridge can be attached to the pick up tube that hangs down in the water bottle by screwing
it into the supplied fitting. To replace, unscrew the old cartridge and screw in the new one. Place the Date
Changed label on the outside of the clean water system as a reminder of when the next change is required.
cAUTiOn:
Do not use the DentaPure DP365 Cartridge for longer than its rated one-year period. The
cartridge will no longer provide disinfection for the system and may plug the unit and obstruct the water flow.
cAUTiOn:
Do not use the DentaPure DP365 if using a shock treatment to disinfect lines as it will damage the cartridge.
The cartridge must be removed and a standard tubing used in the bottle for performing shock disinfection treatment of
the water system.
cAUTiOn:
Do not use bleach to disinfect the water system. Due to its caustic nature, it can corrode certain internal
parts. Any damage is excluded from warranty coverage.
cAUTiOn:
The user assumes all responsibility for maintaining and disinfection of the water system, and for periodically
having water sample tests performed to ensure that disinfection is satisfactory.
There are many other established methods for disinfecting a closed water system that consist of liquids, powders and tablets.
However, the chemicals used in these cleaners may damage the materials of the diaphragms used in the control block and
other valves in the delivery system. Further, tablets may not dissolve completely and may cause blockage of the water system.
Any damage to the system as a result of the use of these alternate cleaners is excluded from warranty coverage.