10.0 Reference lists
10.1 Delivery System Preventative Maintenance Schedules
Category
Time Needed/
Minutes
Frequency
Daily Weekly Monthly Annually
3 yrs
5 yrs or
greater
Delivery Systems
Wipe down work top, and entire delivery system including
holders, with warm water and surface disinfectant
recommended for dental equipment
10
X
perform all manufacturers recommendations for daily
sterilization, lubrication, and maintenance of optional
instruments
Varies
X
Verify all external warning labels are still readable, replace
if needed
10
X
Inspect oil mist recovery jar and replace cotton 2X2 if
saturated
5
X
Inspect hand-piece pressure gauge for correct pSI, replace
if needed pN# 95-0136
5
X
Verify air pressure from main air connection is at 65-80
pSI, adjust if needed.
10
X
Verify all door and drawer hinges and latches are
undamaged, lubricate if necessary.
10
X
Verify operation of water relay valve for opening and
shutting. Replace if necessary
10
X
Replace gasket pucks in control block
90
X
Replace gaskets in handpiece tubing bulkheads
30
X
Replace handpiece tubing if worn or brittle
120
X
Inspect display overlay, replace if needed.
10
X
Foot Pedal
Clean and disinfect pedal and tubing
5
X
Clean and inspect under dome, then lubricate o-rings and
poppet valve pN# 95-0085 (grease)
15
X
Install rebuild kit pN# 95-0240
60
X
Replace water toggle pN# 95-0242
30
X
Replace foot control tubing pN# 22-0016
45
X
Closed Water System
Clean and disinfect closed water system housing
5
X
Inspect closed water system gauge for correct 35 pSI,
replace if needed pN# 95-0266
5
X
Inspect gaskets inside pressure head and replace if torn
pN# 95-0146
10
X
Clean and disinfect pick up tubes pN# 22-0115 and filters
pN# 21-0049, replace if needed
5
X
Replace water bottles pN# 95-0140Qt
5
X
Replace pressure head regulators pN# 95-0142
90
X
Replace 35 pSI inline regulator pN# 95-0216
30
X