2.0 component Operating instructions
Locate the Main Power switch and be sure it is in the On position. This switch must be
turned on in order to operate, and should be turned off at the end of each day.
2.1 clean closed Water System
The Dual Closed Clean Water System (see figure 2.1) features easy-to-fill, pressurized
water bottles that enable you to control the quality of water to your handpieces and air/
water syringe. The water bottle is made of non-porous plastic. The bottle is reusable, but
should be replaced periodically. See maintenance schedule section.
2.1.1 How it Works
Air, regulated to approximately 35 psi, is supplied to the water bottles through tubing
attached to the bottle caps. The air pressure in the bottle forces clean water from the
water bottle into the water pick-up tube and out to the handpiece control blocks, syringe,
and auxiliary quick connects. The water system is a “closed system,” allowing control of water quality and water system
asepsis.
2.1.2 Filling the Bottles
Locate the Dual Closed Clean Water System mounted inside the unit behind the door. Locate the toggle switch for supplying
air pressure to the bottles. Flip the switch to the OFF position to turn off the air supply to the bottles.
Slowly unscrew the bottles to allow them to depressurize. Clean as appropriate. Fill the bottles with clean, demineralized
water to a level below the screw threaded portion. Re-attach the bottles and hand tighten only.
2.1.3 Operating the Dual closed clean Water System
The closed clean water system features a selector valve that allows the operator to choose solution from bottle 1 or bottle 2.
Turn the air pressure toggle switch to the ON position and ensure that the selector switch is flipped to the desired bottle. The
system will now provide solution to the delivery unit and corresponding handpieces.
2.1.4 Disinfecting the Water System
Refer to Cleaning and Disinfection Section of the Manual for disinfection of the clean water system
2.1.5 Purging Water From the System
Running compressed air through the water lines is an easy and effective way to remove water from the lines and prevent
stagnant water from residing in the lines during temporary or long term storage of the system. To purge the water, remove the
bottles and completely empty them of their liquid contents. Then replace the bottles and turn on the air pressure supply to the
bottles. Run the handpiece connections and air water syringe until there is no more water particles exiting them.
2.2 Air/Water Syringe
Check the flow of air, water and spray from the syringe. Press the right button
for water stream. Press the left button for air. Press both simultaneously for
spray. To adjust the flow of air, turn the syringe air adjustment knob clockwise
to increase, counter-clockwise to decrease. To adjust the flow of water, turn the
syringe water adjustment knob until the desired level of flow is achieved (see
Figure 2.2).
2.2.1 Tip Removal and Replacement