4-3
Weekly Care and Cleaning of TriaDyne Proventa Critical Care Therapy System
In addition to daily care and cleaning, the
TriaDyne Proventa
cover sheet should be replaced weekly or more
often as needed. Excessively soiled mattress components and cover sheets should be replaced and laundered
per institution protocol. For replacement mattress components or cover sheets, please contact ArjoHuntleigh.
Preventive Maintenance
It is recommended that all preventive and battery maintenance procedures be performed only by qualified
ArjoHuntleigh service personnel or ArjoHuntleigh approved personnel. Contact ArjoHuntleigh for further
information regarding long-term equipment care and preventive maintenance.
•
Monthly Cleaning
- The cushions should be laundered at least monthly.
•
Air Filter
- Clean / change air filter(s) when necessary.
•
Battery Inspection
- Inspect batteries and replace in pairs, if batteries are more than 28 months old
from the date of manufacture. Install a battery history label on the equipment tray battery cover with
the recorded date when the batteries should next be replaced. Apply label in a visible location so that all
information is displayed.
When not in service, unit must remain plugged in to maintain battery charge.