3-1
Nursing Care
It is recommended that all chapters of this manual be reviewed prior to product use. Carefully read the
Contraindications, Risks and Precautions
and
Safety Information
sections in the
Introduction
chapter
of this manual prior to performing nursing care for a patient on the
TriaDyne Proventa
Critical Care Therapy
System.
CPR
1.
Squeeze and hold red CPR handle located on underside of
head end of unit frame, as shown at right. CPR handles are
located on both sides of unit.
Head end of unit frame (if elevated) will drop abruptly
when CPR handle is squeezed.
• Air is switched off; cushions will deflate.
• Rotation, percussion and pulsation will be
deactivated.
• The message shown at right will appear and flash
in the
Nurse
display panel.
• The display shown at right will appear and flash
on the main control panel.
2.
Lower side rails.
3.
Remove headboard as needed.
4.
Unfasten arm pack strap as needed.
5.
Begin CPR. Crash board is not needed. Patient’s weight and the CPR process will deflate cushions.
6.
After CPR is performed and patient is clinically stable:
• Press
Cancel
CPR
button on main control panel to inflate cushions and resume unit functions.
• Reinstall headboard as needed.
• Refasten arm pack strap as needed.
• Raise and lock side rails. (If used, refer to
Risks and Precautions
in the
Introduction
chapter of this
manual.)
• Resume therapy as directed by physician.
• Adjust patient surface for patient comfort.
CPR
Handle
CPR Activated !
CPR Activated !
Air Switched OFF !
Press to CANCEL CPR>