5-17
5.
Press
Lock
button. The display shown at right will
appear.
• Press
Yes
button if positioning packs
will
be used.
This allows the patient to receive over 20
°
angle of
rotation.
• Press
No
button if positioning packs
will not
be used.
This ensures the patient will not receive over 20
°
angle of rotation regardless of the angle values
entered in the
Rotation
menu.
• The Lock-out menu #1 shown on previous page will appear.
6.
Press
Change
button to adjust the lock-out settings for rotation, pulsation, air adjust and warmer, as
needed.
7.
Press
Next Menu
. The Lock-out menu #2 shown on previous page will appear.
8.
Press
Change
button to adjust the lock-out setting for percussion.
9.
Press
Enter
button to save lock-out settings and then return to the
Home
menu.
Rotation, pulsation, percussion and warmer will be deactivated when the
Lock-Out
menus are
accessed. Only pulsation will be restored when the
Home
menu appears. Reactivate rotation and
percussion from the
Home
menu. Reactivate warmer from the
Warmer
menu.
After lock-outs are set, a beep will sound if a button is pressed to activate an air function that has
been locked-off or frozen. The
Requested Action Not Available
display will appear temporarily.
Accessing
Lock-Out
Menus with Patient on Unit
This method does not require a power-down. Use this method to access the
Lock-Out
menus when a
patient is on the unit.
1.
With
Home
menu present, press and hold the hidden
button for ten seconds.
Diagnostic
menu is displayed.
2.
Press
Lock
button. Refer to steps above in the
Accessing Lock-Out Menus without Patient on
Unit
section to select options in
Lock-Out
menus.
Are You Using the POSITIONING PACKS?
YES>
NO>
Hidden
Button
Содержание Triadyne Proventa