5-29
Nurse Controls on Footend Side Rails
Not all units come equipped with nurse controls on the foot-end side rails.
If equipped, the footend side rail nurse control panels are located on the outside face of the footend side
rails. The
Nurse
display panel is located on the back side of the main control panel.
Right Hold
Use right hold to turn the patient to the target right rotation angle and hold the patient there.
1.
Press
Right Hold
button shown at right.
• The
Turn to Right and Hold
message will
appear in the
Nurse
display panel, as shown
at right.
• The
Right Rotation
display will appear on the main control panel.
• Rotation will be deactivated.
• Pulsation, percussion and warmer will continue, if activated.
INSTAFLATE
Use to increase air pressure in
all cushions and create a firm
surface.
BED POSITIONING FUNCTION
See below
SCALE
Press to display
patient weight in
Nurse
display panel.
SEAT DEFLATE
Use to reduce air
pressure in seat section
cushions by 50%.
Left Hold
Center Hold
Right Hold
On / Off
_____
___
____
____
Turn to RIGHT & HOLD
On / Off
CTR Hold
RT Hold
LT Hold
INSTAFLATE
SEAT DEFLATE
SCALE