Anderson Instrument Company
3-9
AIC2071
Cable Field Preparation
Anderson Instrument Company provides two 25-ft long factory-prepared
cables: Coil Drive and Electrode Signal. If either cable is too long, the excess
cable must be trimmed at the converter end and the wires prepared for
connection in the field. Procedures for adjusting the length of the cable
follow.
Coil Drive Cable
The coil drive cable contains two conductors and a shield wire wrapped in an
aluminum sheath. Make the adjustment to the length of this cable at the
converter end of the cable.
1. Trim the factory prepared cable to the length of cable required. Discard the
factory prepared end.
2. Remove 1.5" of external insulation and foil. Retain the shield wire.
Note:
A factory-supplied 1/4" spade terminal will be crimped on to
the shield wire after the cable has been fed into the converter.
3. Remove .25" of insulation from the black and white wires.
Electrode Signal Cable
The electrode signal cable contains three conductors and a shield wire
wrapped in an aluminum sheath. Make the adjustment to the length of this
cable at the converter end of the cable.
1. Trim the factory prepared cable to the length of cable required. Discard the
factory prepared end.
2. Remove 1.5" of external insulation, fillers, and foil. Trim off the shield
wire entirely.
3. Remove .25" of insulation from the black, white, and red wires.
Содержание IZMSG
Страница 24: ...3 8 IZMSG Compact Electromagnetic Flowmeter AIC2071 Figure 3 8 Meter Body Converter Connections...
Страница 60: ...A 2 IZMSG Compact Electromagnetic Flowmeter AIC 2071 Converter...
Страница 61: ...Anderson Instrument Company A 3 AIC 2071...
Страница 62: ...A 4 IZMSG Compact Electromagnetic Flowmeter AIC 2071 Meter Body...
Страница 63: ...Anderson Instrument Company A 5 AIC 2071...
Страница 79: ......