2-2
IZMSG Compact Electromagnetic Flowmeter
AIC 2071
Caution
Never install the meter body without a
connection seal between the meter body flow tube
and the connection adapter. Never substitute any
other style of seal for the factory-supplied seals.
Location
The meter body provides Anderson measuring results when installed in
either a vertical or horizontal pipe section. Select a location where the meter
body will be completely filled with product. There must be a rise in the
piping at the outlet side of the meter body. Avoid the presence of air, either
bulk or entrained, from entering the product as it passes through the flow
tube.
In a vertical pipe section, install the meter body in the upward direction of
flow. In a horizontal pipe section, install the meter body at the lowest
possible point of the piping system and position the meter body with the axis
of the electrodes horizontal to the ground. This position will reduce sediment
deposits and/or air bubbles on the electrode surface.
Precautions that must be followed when selecting a location for the meter
body include:
P
Do not install the meter body where the ambient temperature
exceeds 130°F.
P
Avoid installing the meter body next to equipment emitting strong
electromagnetic fields that could distort the magnetic field
generated by the flowmeter and cause measuring errors.
P
Do not install the meter body where vacuum conditions may exist.
Содержание IZMSG
Страница 24: ...3 8 IZMSG Compact Electromagnetic Flowmeter AIC2071 Figure 3 8 Meter Body Converter Connections...
Страница 60: ...A 2 IZMSG Compact Electromagnetic Flowmeter AIC 2071 Converter...
Страница 61: ...Anderson Instrument Company A 3 AIC 2071...
Страница 62: ...A 4 IZMSG Compact Electromagnetic Flowmeter AIC 2071 Meter Body...
Страница 63: ...Anderson Instrument Company A 5 AIC 2071...
Страница 79: ......