Anderson Instrument Company
C-1
AIC 2071
Appendix C
IZMSG Technical Specifications
Meter Body
Connections
Sanitary clamp type
Construction
Housing
: AISI Type 304 Stainless Steel
Lining
: PTFE (Non-filled virginTeflon)
Electrodes
: AISI Type 316L Stainless Steel
Connection Box
: Cast aluminum with SBGL corrosion
resistant coating
Entry Wiring Connections
Choice of:
2 water tight cord grips
2 female ½" NPT conduit adapters
Maximum Product Temperature
325 F (160 C)
o
o
Maximum Inlet Pressure
154 psi absolute (11 bar)
Minimum Inlet Pressure
1.4 psi absolute (0.1 bar)
Minimum Fluid Conductivity
Standard
: 100 micromhoes/cm
Special Request
: 5 micromhoes/cm
Magnetic Field
DC pulsed @ 20 Hz with constant current supply
Self correcting zero adjust
Converter
Power Supply
Field selectable by jumpers:
115 VAC/50-60 Hz/single phase
230 VAC/50-60 Hz/single phase
Power Requirements
15 VA (approximately 8 W)
Fuse Protection
315mA, slow response, 5 x 20mm
Scaled Digital Outputs
2 independent, scalable pulsed outputs
Open collector 30 V @ 20 mA
Opto-isolated