2-6
IZMSG Compact Electromagnetic Flowmeter
AIC 2071
Caution
Do not remove the meter body from the pipeline
by disassembling the connection adapters from
the meter body. Connection adapters should be
disassembled and assembled to the meter body
only after the meter body has been removed from
the pipeline.
Converter
The converter is shipped ready for installation. Verify that the converter
serial number matches the meter body serial number. The converter was
factory flow tested and calibrated with the accompanying meter body as a
matched set to the customer's specifications.
Caution
Never operate a meter body and converter with
different serial numbers. The flowmeter will
malfunction or operate improperly.
Location
Install the converter as close as possible to the meter body to avoid
extraneous inductive pick-up. Do not substitute any other cable for a factory-
supplied cable.
The standard factory-supplied cables are 25-feet long. If runs longer than
25 feet are required, contact Anderson Instrument Company. Do not splice
cables. For runs shorter than 25 feet, the cable must be trimmed to the proper
length. See Section 3,
Making Electrical Connections,
for instructions on
field preparation of cables.
Precautions that must be followed when selecting a location for the converter
include:
P
Do not install the converter where the ambient temperature
exceeds 130°F.
P
Do not install the converter where it may be exposed to extreme
vibration or direct sunlight.
Содержание IZMSG
Страница 24: ...3 8 IZMSG Compact Electromagnetic Flowmeter AIC2071 Figure 3 8 Meter Body Converter Connections...
Страница 60: ...A 2 IZMSG Compact Electromagnetic Flowmeter AIC 2071 Converter...
Страница 61: ...Anderson Instrument Company A 3 AIC 2071...
Страница 62: ...A 4 IZMSG Compact Electromagnetic Flowmeter AIC 2071 Meter Body...
Страница 63: ...Anderson Instrument Company A 5 AIC 2071...
Страница 79: ......