3-7
AIC2071
Figure 3-7. Coil Drive Cable Connections at Converter
Cable
Wire
Converter Terminal
Gray - Electrode Signal
Cable
red
14 on X2
white
16 on X2
black
18 on X2
Black - Coil Drive Cable
white
11 on X1
black
12 on X1
shield
see Steps 3 and 4
3. Attach a 1/4 inch crimp connector spade terminal to the coil drive cable
shield wire.
4. Loosen Main Board attachment lug screw located in the lower left corner of
the converter. Slide the spade terminal between the Main Board and
underside of the head of the lug screw. Retighten the lug screw.
Caution
Before arc welding on a pipeline with a flow m eter
installed, disconnect the electrode signal cable at
term inals 13, 14, 16, and 18 at the meter body or
term inals 14, 16, and 18 at the converter.