Anderson Instrument Company
5-17
AIC 2071
Table 5-6. Review Checksums (Menu 06)
Parameter
Name
Menu 06
Review Checksums
Remark
eprsum
Display of checksum of the firmware
program
Read-only value is based on the
program version
eparasum
Display of checksum for the critical
calibration parameters. (Parameters that
cannot be unlocked by Function 98.)
Read-only value is based on the
adjusted parameter data
nparasum
Display of checksum for non-critical
parameters. (Parameters that can be
unlocked by Function 98.)
Read-only value is based on the
adjusted parameter data
sentype
Display of meter body type
Read-only value
senno.
Display of meter body number
Read-only value
switch position
Shows position of CAL/MEAS switch
user
Display software version
Read-only value
System 55
CS3 bus communication software version
Read-only value
Appli 5A
Flow measuring software version
Read-only value