Parameter
Name
Menu 02
Meter Parameters
Remark/Range/Default
Anderson Instrument Company
5-13
AIC 2071
standby mode
Reduces power consumption during metering
pauses
by turning off meter coil excitation current
standby mode 0:
The standby function can be
activated by pressing the “F” key followed by “99"
or by activating the digital input
IN1
standby mode 1:
standby not possible
standby mode 2:
standby is possible
Range:
0 - 2
Default:
1
parameter mode
Security function
for the parameters, which
allows the parameters to be protected against
unauthorized changes
0
ÿ
the calibration switch
"CAL/MEAS"
locks only the critical calibration
parameters
1
ÿ
the
"CAL/MEAS"
switch locks all
parameters
2
ÿ
allows operator to unlock non-critical
parameters by pressing the “F” key
followed by “98"
Range:
0 - 2
Default:
1