![Anderson IZMSG Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/anderson/izmsg/izmsg_instruction-manual_2941751020.webp)
3-4
IZMSG Compact Electromagnetic Flowmeter
AIC2071
Figure 3-5. Electrode Signal Cable
Figure 3-6. Coil Drive Cable
Connection cables can be routed through a rigid, dedicated conduit using
water tight connectors at the meter body and converter enclosure. If desired,
both cables may be routed in the same conduit. Separate the cables into two
individual conduits adjacent to the meter body before routing the cables into
the field terminal box on the meter body.
Caution
Never bundle or route either the electrode signal
or the coil drive cable near any AC power line.