2 line of the display
Function
nd
D11 D14 D17 D20 D5 D6 D7 D8
Anderson Instrument Company
A-13
AIC 2071
Curr1 on
coil current is constantly at app. +100 ... 110 mA metered at
terminal #11/#12
Curr2 on
value of the metered coil current must be almost the same as “
Curr1
on
” (±0.3 mA) but with the reverse polarity
end hardware test
Quit with key ENTER
use the key
to end this function
x = LED on
o = LED off