3-26 | Thermox Series 2000 WDG-IV / IVC / IVCM
standard sensor connections
All WDG sensors, regardless of the sensor type, require that you make
the following connections from the control unit to the sensor (see Figure
3-27):
Control Unit label
sensor terminal number
Cell +
1
(one pair)
Cell -
2
Furnace
3
(one pair)
15V COM
4
T/C +
5
(one pair)
T/C -
6
15V SPLY
12
(one pair)
15V SPLY
12
Combustibles connections
If you have a combustibles detector in your sensor, also make the fol-
lowing connections between the control unit wiring card and the sensor
board on the sensor (H
2
represents the hydrogen/carbon monoxide style
of combustibles detector. This detector can measure both CO and H
2
) -
refer to Figure 3-27:
Control Unit label
sensor terminal number
H2 Act +
7
(one pair)
H2 Act -
8
H2 REF +
9
(one pair)
H2 REF -
10
Methane connections
If you have a methane detector, also make the following connections (see
Figure 3-27):
Control Unit label
sensor terminal number
CH4 ACT +
13
(one pair)
CH4 ACT -
14
CH4 REF +
15
(one pair)
CH4 REF -
16
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...