Controller / User Interface | 4-13
Communication
Define RS-485 communications parameters between the Series 2000 con-
trol unit and a host device. This includes the following:
Baud Rate
Allows you to define the baud rate that the control unit and the
host computer are using to communicate. Choices are 300,
600, 1200, 2400, 4800, and 9600 baud.
Node Address
Allows you to define the physical address for the control unit
(between 0 and 255) on a network so the host computer can
identify the control unit if using network communications (see
below). Note that each control unit must be assigned a unique
address; otherwise, communication results will be unpredict-
able.
Enable Serial
Allows you to temporarily disable control unit serial communi-
cations.
Combustibles Option
Combustibles
Scale
Choices are % combustibles or PPM combustibles.
The % combustibles full scale range is 0 to 5 %.
The PPM combustibles full scale range is 0 - 10,000 PPM.
If you have the methane option, the range is always in % methane (not select-
able), and the full scale range is 0 to 5 %.
If you have a methane detector, the combustibles detector must be used in the
PPM range.
Combustibles
Type
Choices are Hot Wire and RTD detector.
If not sure, check your customer configuration sheet, in
-
cluded in the front of this manual.
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...