A-14 | Thermox Series 2000 WDG-IV /I VC / IVCM
12/18
I1_span
Analog output 1, 20 ma value
R/W
13/19
I1_zero
Analog output 1, 0 or 4 ma value
R/W
14/20
I2_span
Analog output 2, 20 ma value
R/W
15/21
I2_zero
Analog output 2 , 0 or 4 ma value
R/W
16/22
I3_span
Analog output 3 , 20 ma value
R/W
17/23
I3_zero
Analog output 3 , 0 or 4 ma value
R/W
18/24
I4_span
Analog output 4 , 20 ma value
R/W
19/25
I4_zero
Analog output 4 , 0 or 4 ma value
R/W
1A/26
I_filt1
Filter value for analog output 1
R/W
1B/27
I_filt2
Filter value for analog output 2
R/W
1C/28
I_filt3
Filter value for analog output 3
R/W
1D/29
I_filt4
Filter value for analog output 4
R/W
1E/30
Alm3_value
alarm setpoint for process alarm
IQ
Oxygen alarm setpoint for Alarm #3
S2
Value
is always in percent
R/W
1F/31
Alm4_value
alarm setpoint for Alarm #2
IQ
Oxygen alarm setpoint for Alarm #4
S2
Value
is always in percent
R/W
20/32
Alm5_value
Combustibles PPM alarm #5 setpoint
S2
Alarm #3 setpoint
IQ
R/W
21/33
Alm6_value
Combustibles PPM alarm #6 setpoint
S2
R/W
22/34
Alm5_percent
Combustibles percent alarm #5 setpoint
S2
R/W
23/35
Alm6_percent
Combustibles percent alarm #6 setpoint
S2
R/W
24/36
Alm7_percent
Methane percent alarm #7 setpoint
S2
R/W
25/37
Alm8_percent
Methane percent alarm #8 setpoint
S2
R/W
26/38
Gas1_time
Amount of time oxygen span gas will flow in
seconds
IQ
Amount of time oxygen span gas will flow in
MMSS
S2
R/W
27/39
Gas2_time
Amount of time oxygen zero gas will flow in
seconds
IQ
Amount of time oxygen zero gas will flow in
MMSS
S2
R/W
28/40
Gas3_time
Amount of time the combustibles span gas
will flow in seconds
IQ
Amount of time combustibles span gas will
flow in MMSS
S2
R/W
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...