Calibration | 5-21
Set Cal timers - Set auto timers
Set the frequencies for automatic calibrations or verifications, including
the time when the first calibration or verification should begin. You have
the option of setting the cycle in either hours or days.
If you schedule a verification and calibration at the same time, the
verification will be performed first.
1. From the
Set Cal Timers
menu, select
Set Auto Timers
.
2. Select
Auto Cal Cycle
if defining an automatic calibration cycle. Select
Auto Verify Cycle
if defining an automatic verification cycle.
3. Select
Set Cycle in DAYS
or
Set Cycle in HOURS
, depending on
whether you want to set the frequency of calibrations or verifications
in days or hours.
If you are setting the auto timers for the first time, the Series 2000
software will automatically prompt you to enter the current time and
date.
You will be prompted to enter the frequency for the calibration or verifica-
tion cycle (either in days or hours, depending on what you selected:
Cal Cycle = 01 days
New Value = xx
4. Enter the number of days (or hours) for the automatic calibration or
verify cycle. As you enter the values, they overwrite the XXs on the
display. To enter 1 day or 1 hour, enter 01; otherwise, the system will
assume you are entering 10 days (or hours).
You will be prompted to enter the time of day when the first calibra-
tion of the cycle should begin:
Cal time is YY:YY
Enter new time Hr:Mn
5. Enter the time you want the first calibration or verification to begin,
then press
Enter
. The system uses military time, so to enter 4:00 p.m.,
you would type 16:00. If, for example, you set the cycle frequency at
8 hours, and specify the first calibration to take place at 4:00 p.m., the
next automatic calibration or verification will take place at midnight.
NOTE
NOTE
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...