Calibration | 5-19
Combustibles and Methane option
If you have the combustibles option, you will be prompted to press
Enter
when you have set the pressure and flow for the zero gas:
ENTER for next Gas.
CANCEL to quit.
5. When you have adjusted the pressure and flow for the zero gas to the
desired levels, press
Enter
.
The following message will appear on your display, indicating that the
RCU has introduced the combustibles span gas to the sensor:
Comb gas is flowing
Once you have set the pressure and flow for the span gas, the system
prompts you to press the
Cancel
key to end the cycle:
CANCEL to quit
If you have the methane option, you will be prompted to press
Enter
when you have set the pressure and flow for the zero gas:
ENTER for next Gas.
CANCEL to quit.
6. When you have adjusted the pressure and flow for the zero gas to the
desired levels, press
Enter
.
The following message will appear on your display, indicating that the
RCU has introduced the combustibles span gas to the sensor:
Methane gas is flowing
Once you have set the pressure and flow for the span gas, the system
prompts you to press the
Cancel
key to end the cycle:
CANCEL to quit
7. Press the
Cancel
key to end the cycle.
Then, if you specified a recovery time, the system will display the recov-
ery time remaining:
Recovery xx:xx
The system returns to monitoring process readings.
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...