7-10 | Thermox Series 2000 WDG-IV / IVC / IVCM
Combustibles Detector Replacement
Always use a backup wrench when working on sensor plumbing.
Hot-Wire Combustibles Detector Replacement
Use Figure 7-5:
1. WDG-IVC
Remove the detector leads from terminal strips E, F, H, and I on the
sensor board.
WDG-IVCM
Remove the detector leads from terminal strips W, X, Y, and Z.
2. Loosen the compression fitting nut that holds the combustible detector
in place and remove the detector from its housing. Save the nut from
the detector to use when installing the new combustibles detector.
3. It is
very important
to orient the new combustibles detector correctly
inside the combustibles detector block (sample flow path must contact
both detector leads at the same time).
Place a mark on the edge of the combustibles detector at the 12 o’clock
position after you have determined the correct orientation (Figure
7-5). Place detector elements in detector block vertically.
4. Install the new combustibles detector with nut and ferrules. Be ex-
tremely careful not to damage the detector elements when installing.
Ensure that the mark you made in Step 3 is in the 12 o’clock position
to ensure proper orientation of the combustibles detector in the com-
bustibles detector block.
5. Reconnect the detector leads to the sensor board.
The active element wires are labeled
“A” and “AC”
.
WDG-IVC
connect to terminals E and F on the sensor board.
Polarity not required.
WDG-IVCM
Connect to terminals W and X.
The reference element wires are labeled
“R” and “RC”
WDG-IVC
Connect to terminals H and I on the sensor board.
Polarity not required.
WDG-IVCM
Connect to Y and Z
NOTE
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...