7-8 | Thermox Series 2000 WDG-IV / IVC / IVCM
Furnace Replacement
Disconnect power from the sensor and control unit and allow unit to
cool before replacing parts inside the sensor.
Use Figure 7-4:
1. Disconnect cell clips.
2. Remove the thermocouple from the furnace (see the “Thermocouple
Replacement” section in this chapter for help).
3. Disconnect the furnace leads from terminals R and S on the sensor
board.
4. Loosen the compression fitting on the underside of the cell housing
and the union fitting on the convection loop return. To prevent strip-
ping threads on any of the fittings, we suggest that you loosen each in
turn at small increments - don’t completely loosen the one fitting and
then try to loosen the other fitting.
5. While holding the furnace in one hand, remove the convection loop
mounting clamp. Be sure to save the mounting clamp for when you
install the new furnace.
6. Gently pull the cell housing out of the furnace and remove the fur-
nace.
7. Slip the new furnace onto the cell housing and attach the cell housing
using the convection loop mounting clamp.
8. Remount the cell housing (reverse the actions taken in Step 4).
9. Reinstall the thermocouple (reverse the actions taken in Step 2).
10. Using an Ohm meter, check for short circuits of the heater coil to the
metal thermocouple sheath. Check that neither furnace lead is short
circuited to the metal furnace cover. The resistance should be infinite.
If you measure a short circuit, loosen the screw that secures the ther-
mocouple and relocate slightly until resistance is infinite.
11. Reattach cell clips.
12. Reattach furnace leads (reverse the actions taken in Step 3).
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...