4-24 | Thermox Series 2000 WDG-IV / IVC / IVCM
You are prompted to enter the Alarm 8 set point value:
Alarm #8=xx.x % CH4
New Value?
7. Enter the Alarm 8 set point value. Note that methane alarms are al-
ways high alarms. So, for example, Alarm 7 could be a high alarm, and
Alarm 8 high/high.
8. Press
Cancel
or
Enter
to exit this menu option.
Alarm Configure
Define whether alarms should trigger based on a high alarm or a low
alarm condition. If you have the combustibles or methane options, Alarms
5 through 8 always act as high alarms. So, for example, Alarm 5 could be a
high combustibles alarm, and Alarm 6 could be a high/high combustibles
alarm.
Whether Alarm 3 is set high or low only applies if it is set to track
oxygen levels (% O
2
). If Alarm 3 is set to trigger at the start of a cali-
bration or verification, this option does not apply.
Configuring the alarms
1. From the
Alarm
menu, select
Alarm Configure
.
2. Define Alarm 3 and Alarm 4 as high or low alarms.
3. Press
Cancel
to exit this menu option.
Relay Configure
All Series 2000 control unit relays are Form A, normally open contacts.
You can use the software to set whether these alarms should close on
an alarm condition (energize on alarm) or open on an alarm condition
(de-energize on alarm). This menu option applies to all alarms except for
the service and watchdog alarms, which are always set to de-energize on
alarm (fail-safe).
NOTE
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...