4-16 | Thermox Series 2000 WDG-IV / IVC / IVCM
You will be prompted to define the new value to correspond to the 20 mA
analog output:
Out # y 20 ma = xx
New Value?
where y is the analog output port, and xx is the present value that the cur-
rent output currently represents (for example, 10 % O
2
). The output port
that appears depends on the analog port you selected.
3. Input the value to correspond to the 20 mA analog output.
You will be prompted to define the value to correspond to the 0 or 4 mA
analog output:
Out # 1 4 ma = xx
New Value?
where
xx
is the value that the current output currently represents.
4. Input the value to correspond to the 4 (or 0) mA analog output.
Set Track/Hold Option
Define the signals to send to the selected analog output port during a cali-
bration or verification. You have the option to either hold the last process
reading or track calibration readings. Separate decisions can be made for
verify and calibrate operations, so you can hold the last process readings
during verifications, but track calibration gas readings during calibrations.
Setting track or hold
1. Select an analog output port.
2. From the
Analog Range
key, select
Set Track/Hold.
You can choose to either
hold
process readings or
track
calibration
readings on the analog output port. Separate choices can be made for
calibration and verify operations.
For
calibration
operations, select one of the following:
Hold during cal
or
Track during cal
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...