Installation | 3-23
SHIELD RING METHOD
GROUND STUD METHOD
Shield Terminal Ring
Conduit Housing
Cable Shield(s)
Max. Length 1 inch
Conduit Nut
Quick Disconnet (1/4”)
Capacitor
Conduit Nut
Cable Shield(s)
Max. Length 2 inches
Nut
Washer
Ring Terminal
Conduit Housing
Capacitor
Figure 3-18. Direct shield grounding methods
transient and RFI interference
This section describes transient and RFI interference precautions:
•
Although there are transient and noise protectors on all control unit
I/O connections (communications, current outputs, sensor, etc.), this
protection is intended to act as a last line of defense against unwanted
transient and RFI interference.
•
Proper installation practices to prevent the introduction of transients
and noise into the system must be followed. Inductive loads connect-
ed to the control unit must have transient suppressors installed at the
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...