4-14 | Thermox Series 2000 WDG-IV / IVC / IVCM
Analog Range Key
Set Current
Range
Set Track/Hold
Set Current
Mode
Output Filtering
Select Function
ANALOG RANGE KEY
Figure 4-3. Analog Range Key menu.
You can direct process readings to either of the two analog output ports
on the rear of the control unit. If you have the combustibles option, you
can also have process readings directed to two additional analog output
ports activated with this option. All functions in this chapter are directed
to a specific analog output port.
ANALOG KEY
FUNCTIONS/OPTIONS
Set Current Range
Define what information should be sent to the analog
output port during a calibration. For example:
4% O
2
= 4 mA
10% O
2
= 20 mA
Set Track/Hold
Hold output at last process reading during a cal or
verify, or track the calibration readings.
Set Current Mode
Define analog output range:
0 to 20 mA
4 to 20 mA
Output Filtering
Smooth out variations in readings
Select Function
Define the information the analog output should track:
% O
2
Cell temperature
Cell Millivolts
Thermocouple Millivolts
Combustibles (option)
Methane (option)
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...