5-12 | Thermox Series 2000 WDG-IV / IVC / IVCM
Calibration with RTD Type Combustibles Detector
If you are calibrating a 3-in-1 sensor with the RTD combustibles de-
tector, your combustibles span calibration gas must include methane.
On RTD-type detector
, the zero gas display then appears:
O2 xx.x % Set xx.x %
Cmb yyyy Set 0
CH4 zzzz Set 0
Press ENTER to Zero
Display line 2 will only appear if you have the combustibles option;
display line 3 will only appear if you have the combined combustibles
and methane option.
1. Press
Enter
when the zero gas reading(s) on the left of the display stabi
-
lizes. The reading on the far right of the first display line shows the setpoint
value you entered for the O
2
zero gas using the
Cal Gas Values
menu
option. This O
2
zero gas setpoint value should match the cal gas cylinder
for the O
2
zero gas.
If you have the Comb. & CH4 span gas option, you are also prompted
to apply the Comb. & CH4 span gas to the calibration gas inlet port of
the sensor (tubing must be free of oil and dirt):
Inject Comb Span Gas
ENTER to Continue
CANCEL to Abort
2.
Press
Enter
when you have applied this calibration gas. You are prompted
to calibrate the Comb & CH4 span gas.
O2 xx.x %
Cmb yyyy Set yyyy
CH4 zzzz Set zzzz
Press ENTER to Span Comb
3.
Press
Enter
when the Comb & CH4 span gas reading(s) stabilizes on the
far left of the display.
NOTE
NOTE
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...