5-16 | Thermox Series 2000 WDG-IV / IVC / IVCM
Cal Gas Value
Enter span and zero calibration gas values to match the calibration gas
cylinders you use to calibrate or verify your system. These values will be
used for all automatic, remote, or manual calibrations and verifications.
•
O
2
span gas - high calibration gas
•
Zero gas - low calibration gas
•
Combustibles span gas (optional)
•
Methane span gas (optional)
The O
2
span gas value you enter must be higher than the O
2
zero gas
value.
1. From the
Calibrate
key menu, select
Cal Gas Value
.
You are prompted to enter the span gas value:
Span Gas = xx.xx% O2
New Value?
2. Enter the span gas value, then press
Enter
. You can also include a
decimal point (for example, 20.9% O
2
). If you want to keep the current
span gas value unchanged, press
Enter
without entering a value.
You are prompted to enter the zero gas value:
Zero Gas = xx.xx% O2
New Value?
3. Enter the zero gas value, then press
Enter
. You can enter up to two
decimal points (for example, 2.00% O
2
). If you want to keep the cur-
rent zero gas value unchanged, press
Enter
without entering a value.
Combustibles option
If you have the combustibles option, you are then prompted to enter
the combustibles span gas value:
Comb Span = xxxx YY
New Value?
NOTE
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...