2-6 | Thermox Series 2000 WDG-IV / IVC / IVCM
Sensor : WDG-IVC - 0-500 PPM
Output Range
From 0-1% to 0-100%
Oxygen
Combustibles
From 0-1% to 0-100%
From 0-500 ppm to 0-10,000 ppm or from 0-1% to 0-5%
Accuracy
Oxygen
± 0.75% of measured value or ± 0.05% oxygen, whichever is greater
Combustibles
± 25 ppm between 0 to 500 ppm or ±5% of full scale output range
Response O2
90% of a process step change < 30 secs.
Aspirator Air
Requirements
10 to 20 scfh (4.72 to 9.4 L/min) at 15 to 100 psig (1.05 to 7.04 kg/cm
2
)
Max. Flue Gas Temp / Probe Type / Lengths
1300°F (704°c)
1875°F (1024°c)
3000°F (1649°C)
316 ss
310 ss
Hexoloy
®
36” - 108” (91 cm - 271 cm)
36” - 108” (91 cm - 271 cm)
24” - 72” (60 cm-182 cm)
Max. Sample
Dewpoint
392°F (200°C)
Sample Pressure
± 10 inch water gauge
Environment
Ambient Temp.:
-5°F to 160°F (-20°C to 71°C);
-5°F to 140°F (-20°C to 60°C) with Div. 2 Option
Relative Humidity:
10% to 90%, non-condensing
Enclosure
Hinged, stainless steel, weatherproof NEMA 4X (IP56). Flame arrestors standard. Optional
CENELEC and NEC Class I, Division 2 available.
Power Requirements
115 VAC, ± 10%, 47-63 Hz, 600VA max.
(650 VA max. floor mount option)
230 VAC, ± 10%, 47-63 Hz, 1850 VA max.
(1900 VA max. floor mount option)
Calibration Gas Requirements
Use calibration gases @ 10 PSIG, 1.5 SCFH (0.70 kg/cm
2
, 0.7 L/min.)
o
2
span gas:
Air or from 1.0 to 100% o
2
, balance n
2
o
2
, Zero & comb
Zero gases:
2% or from 0.1 to 10% o
2
, balance n
2
comb span gas:
80 to 100% of the selected combustibles recorder output range in certi-
fied mixtures of CO + H
2
, 3-4% o
2
, balance n
2
System Compliance
EMc compliance:
Safety Compliance:
2004/108/Ec
73/23/EEc
NOTES
1.
All static performance characteristics are with operating variables constant.
2.
System accuracy referenced to 0.1 to 10% calibrated range.
3.
Response is to calibration gas (with flame arrestors)
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...