Serial Communications
|
A-19
61/97
Alarm_flags
IQ
Which oxygen alarm has been set and why
bit 0
– high oxygen, process alarm
bit 1
– low oxygen, process alarm
bit 2
– high combustible, process alarm
bit 3
– high methane, process alarm
bit 4
– high oxygen, alarm #2
bit 5
– low oxygen, alarm #2
bit 6
– high combustible, alarm #2
bit 7
– high methane, alarm #2
bit 8
– high oxygen, alarm #3
bit 9
– low oxygen, alarm #3
bit 10
– high combustible, alarm #3
bit 11
– high methane, alarm #3
bit 12
– Future use
bit 13
- Out of Range, for SA/F sensor
bit 14
- Future use
bit 15
- Future use
Read only
Alarm_flags
S2
Which oxygen alarm has been set and why
bit 0
– low oxygen, alarm #3
bit 1
– high oxygen, alarm #3
bit 2
– low oxygen, alarm #4
bit 3
– high oxygen, alarm #4
bit 4
– high combustion, alarm #5
bit 5
– high combustion, alarm #6
bit 6
– high methane, alarm #7
bit 7
– high methane, alarm #8
bit 8
– Future use
bit 9
– Future use
bit 10
– Future use
bit 11
– Future use
bit 12
– Future use
bit 13
- Future use
bit 14
- Future use
bit 15
- Future use
Read only
Содержание Thermox Series 2000 WDG-IV
Страница 14: ...xiv Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 32: ...2 8 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 59: ...Installation 3 27 Figure 3 20 Sensor connections with all sensor options ...
Страница 67: ...Installation 3 35 Figure 3 24 Series 2000 2 wire RS 485 communication connections ...
Страница 69: ...Installation 3 37 Figure 3 25 Series 2000 4 wire RS 485 communication connections ...
Страница 72: ...3 40 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 100: ...4 28 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 126: ...5 26 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 143: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor ...
Страница 145: ...Service and Parts 7 5 Figure 7 2 Cell replacement ...
Страница 155: ...7 15 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 181: ...A 26 Thermox Series 2000 WDG IV I VC IVCM This page intentionally left blank ...
Страница 185: ...B 4 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...
Страница 191: ...D 2 Thermox Series 2000 WDG IV IVC IVCM This page intentionally left blank ...