GUI reference
R&S
®
ZND
397
User Manual 1173.9557.02 ─ 61
Detect Port Assignment
Starts a procedure by which the R&S
ZND (with a little help from the attached calibra-
tion unit) auto-detects the connected ports. The automatic assignment replaces the
configured one.
In case auto-detection fails
●
an error report is shown as a warning dialog
●
the undetected port connections are marked with warning signs
●
the calibration can be invalid
Note:
For a unidirectional R&S
on page 80) auto-detection at VNA port 2 does not work.
Remote command:
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
Prev/Next
Navigates between the port assignments. Only available if multiple port assignments
are required.
Apply/Cancel
Apply the calculated system error correction to the active channel (or to all channels in
the active recall set, if all channels were calibrated).
The Apply button is enabled when calibration sweeps have been successfully per-
formed for all required port assignments.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:SAVE
Extended cal unit settings:
MMEMory:AKAL:FACTory:CONVersion
MMEMory:AKAL:USER:CONVersion
SYSTem:COMMunicate:AKAL:CONNection
SYSTem:COMMunicate:AKAL:MMEMory[:STATe]
5.11.1.4
Calibration Setting wizard
The "Calibration Setting" wizard guides you through the setup and execution of a man-
ual system error correction.
Cal softtool