GUI reference
R&S
®
ZND
497
User Manual 1173.9557.02 ─ 61
Remote command:
CALCulate:FMODel:DELT:SWEep:CONTrol
CALCulate:FMODel:DELT:SWEep:CONTrol:STEP?
CALCulate:FMODel:DELT:SWEep:CONTrol:IFBW?
Sweep Frequency
Defines the sweep range of the Delta-L channel. The entered "Minimum" and "Maxi-
mum" values are set as
when the Delta-L settings are applied
using "OK". Furthermore, the VNA collects data for the LPI computation a the "Mini-
mum" and "Maximum" frequencies (in addition to the other specified
The default "Minimum" is the standard Delta-L start frequency of 10
MHz. The default
"Maximum" is the instrument's maximum frequency, which can also be set using the
"Set to Max Frequency" button.
If you set the "Minimum" to a non-standard value, a popup indicates that the Delta-L
measurement possibly does not provide accurate results. Same is true for the "Maxi-
mum", which is inevitable, because the standard Delta-L maximum frequencies of
20
GHz and 40
GHz are far beyond the frequency range of the R&S
ZND.
Remote command:
CALCulate:FMODel:DELT:SWEep:FREQuency:MINimum
CALCulate:FMODel:DELT:SWEep:FREQuency:MAXimum
Trace
On
, the VNA deletes the existing traces and sets up the traces whose measure-
ment is required for the selected Delta-L
If
"Use Recommended Trace Configuration"
is checked, then each of the new
traces is displayed in a separate diagram. Otherwise, the VNA places all traces in a
single diagram.
ZND-K2 is installed,
"Show TDR Impedan-
ces"
adds the TDR-based converted impedance traces to the recommended trace
configuration.
Remote command:
CALCulate:FMODel:DELT:TCONfig
CALCulate:FMODel:DELT:TDR
OK
"OK" applies the Delta-L settings and starts the
5.13.11.2
Delta-L measurement
When the Delta-L measurement starts, the analyzer firmware prepares the Delta-L
measurement channel, populates the diagram area with the required measurement
traces, and opens the
Delta-L Measurement dock widget
Delta-L Measurement dock widget
The "Delta-L Measurement" dock widget is where coupon measurements are per-
formed and/or corresponding touchstone files are selected.
Offset Embed softtool
www.allice.de