Command reference
R&S
®
ZND
883
User Manual 1173.9557.02 ─ 61
<CalGroupFile>
Optional string parameter to specify the name of the cal group
file. Cal group files must have the extension
*.cal
. The direc-
tory must not be specified; the analyzer always uses the default
path
C:\Users\Public\Documents\Rohde-Schwarz\Vna
\Calibration\Data
. If there is no link between
<Ch>
and the
specified file, the command has no effect. If no file is specified,
the command resolves any link between
<Ch>
and an arbitrary
cal group file.
Example:
Usage:
Setting only
Manual operation:
"Resolve Pool Link / Remove Pool Link"
MMEMory:LOAD:CORRection:TCOefficient<Ch>
<TraceFile>[, <Trace>[,
<DeEmbedding Position>]]
Loads two-port transmission coefficients from the specified power meter correction file
or trace file to channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)
BOTH
Both positions
(parameter omitted)
If the parameter is omitted, the command refers to the two-port
selected using
SOURce<Ch>:POWer<PhyPt>:CORRection:
Setting parameters:
<TraceFile>
String parameter specifying the name and directory of the loa-
ded trace file. The R&S
ZND supports power meter correction
list files (
*.pmcl
*csv
, and Touchstone
(
*.s1p, *.s2p, ...
) files.
If no path is specified the analyzer searches the current direc-
tory, to be queried with
The file extensions
*.s<n>p
,
*csv
, and
*.pmcl
for Touch-
stone, ASCII, and power meter correction list files are manda-
tory.
*RST:
n/a
SCPI command reference
www.allice.de
Allice Messtechnik GmbH