Command reference
R&S
®
ZND
1031
User Manual 1173.9557.02 ─ 61
Example:
RST; SOUR:POW:CORR:TCO:DEF 1GHz, -5, DUT; DEF
2GHz, -10, DUT; DEF2?
Define two points in the power loss list of the two-port in front of
the DUT; query the second point.
SOUR:POW:CORR:TCO:INS2 1.5 GHz, -7.5, DUT
Insert a third point as point no. 2 in the list.
SOUR:POW:CORR:TCO:COUN? DUT
Query the number of points. The response is
3
.
SOUR:POW:CORR:TCO ON, DUT
Enable the use of two-port transmission coefficients.
SOUR:POW:CORR:TCO:FEED 'Trc1', DUT
Replace the previous 3 points by the trace points of the default
trace "Trc1".
MMEM:STOR:CORR:TCO 'C:
\Users\Public\Documents\Rohde-Schwarz\Vna
\PowerMeterCorr\Test.pmcl', DUT
Store the power loss list to the power meter correction list file
Test.pmcl
in the directory
C:\Users\Public\Documents\Rohde-Schwarz\Vna
\PowerMeterCorr
.
MMEM:LOAD:CORR:TCO 'C:\Users\Public\Documents
\Rohde-Schwarz\Vna\PowerMeterCorr\Test.pmcl',
DUT
Re-load the stored power meter correction list file.
SOUR:POW:CORR:TCO:DEL1, DUT
Delete the first point in the list.
SOUR:POW:CORR:TCO:DEL:ALL, DUT; :SOUR:POW:CORR:
TCO? DUT
Clear the entire list. Query whether the transmission coefficients
are still taken into account. The response is
0
.
MMEM:LOAD:CORR:TCO 'C:\Users\Public\Documents
\Rohde-Schwarz\Vna\Traces\Test.s2p', 'S21', DUT
Load a power loss list from a previously created 2-port Touch-
stone file.
Manual operation:
See
7.3.15.5
SOURce:POWer:CORRection... (other)
The
SOURce:POWer:CORRection...
commands control a scalar source power cali-
bration.
SOURce<Ch>:POWer:CORRection[:ACQuire]
................................................................1032
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]:VERification:RESult?
SOURce<Ch>:POWer:CORRection:DATA
...................................................................... 1033
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
.......................................... 1034
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?
SOURce:POWer:CORRection:NREadings
......................................................................1036
SOURce<Ch>:POWer:CORRection:OSOurces[:STATe]
................................................... 1036
SOURce:POWer:CORRection:PMETer:ID
...................................................................... 1036
SCPI command reference
www.allice.de
Allice Messtechnik GmbH