Command reference
R&S
®
ZND
1034
User Manual 1173.9557.02 ─ 61
Suffix:
<Ch>
.
Calibrated channel number
Parameters:
<Source>
Identifier for the source of the calibrated wave:
'
A<n>
' or '
PORT<n>
' denote correction data for the analyzer port
no.
<n>
.
<CorrData>
Power correction values either in ASCII or block data format,
depending on the current
All numbers are interpreted as dB-values.
Example:
*RST; :SWE:POIN 10
Reset the instrument and reduce the number of sweep points to
10.
SOUR:POW:CORR:ACQ PORT,1
Perform a source power calibration using port 1 as a source
port.
SOUR:POW:CORR:DATA? 'A1'
Query the correction values. The analyzer returns 10 comma-
separated real numbers.
SOUR:POW:CORR:DATA 'A1', 1, 2, 3, 4, 5, -6, -7,
-8, -9, -0
Replace the correction values by ten (new) numbers.
Manual operation:
See
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
Gets settings of the available power calibrations.
Suffix:
<Ch>
.
Number of the calibrated channel.
<Wv>
Number of the power calibration.
Use
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:
on page 1035 to query the number
of available power calibrations.
Query parameters:
SettingType
WAVE | STARt | STOP | POINts | STYPe | ATTenuation |
CPOWer | CFRequency | TSTamp | LTSTamp | FSMode
WAVE
Calibrated port and calibration type (Gen=Source Power,
Rec=Measurement Receiver)
STARt
Start Frequency or Power
Stop Frequency or Power
POINts
Nr. of sweep points
SCPI command reference
www.allice.de
Allice Messtechnik GmbH