Command reference
R&S
®
ZND
1032
User Manual 1173.9557.02 ─ 61
SOURce:POWer:CORRection:PPOWer
......................................................................... 1037
SOURce:POWer:CORRection:PSELect
......................................................................... 1037
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe
........................................................ 1037
SOURce<Ch>:POWer:CORRection[:ACQuire]
<SourceType>[, <SourcePort>[,
<CalOnlyPortFreq>]]
Selects the source for the source power calibration, starts and applies the source
power calibration.
Note:
The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Suffix:
<Ch>
.
Calibrated channel number
Setting parameters:
<SourceType>
PORT | A1 | A2 | A3 | A4
PORT
–
analyzer port is the source of the calibrated wave.
The port number is specified by means of the
<SourcePort>
parameter.
A1 | A2 | A3 | A4
–
direct analyzer port selection. The parame-
ters A1 and A2 are ZVR-compatible. No additional port number
parameter needs to be specified.
<SourcePort>
Analyzer port number, if PORT is the first parameter.
<CalOnlyPortFreq>
Relevant for frequency conversion measurements only.
OFF | 0
The reference receiver at port <CalPort> is calibrated for all fre-
quencies that are relevant for any of the ports.
ON | 1
Only those frequencies are calibrated that are relevant for this
port. This results in shorter calibration times.
*RST:
0
Example:
SOUR:POW:CORR:OSO:STAT OFF
To improve the accuracy, switch off all other sources during the
calibration sweep.
SOUR:POW:CORR:NRE 2; COLL:AVER:NTOL 0.5
Increase the number of readings and reduce the power toler-
ance to improve the accuracy.
SOURce:POW:CORR:COLL:CFAC 0.9
Reduce the correction factor to 0.9.
SOURce:POW:CORR:PSELect PPOW; PPOWer -5
Define reference receiver cal power which is independent of the
port power; set its value to -5 dBm.
SOUR:POW:CORR:ACQ PORT,1
Perform a source power calibration using port 1 as a source
port.
SOUR:POW:CORR:STAT?
Check whether the calibration is applied (the response is
1
).
SCPI command reference