Command reference
R&S
®
ZND
972
User Manual 1173.9557.02 ─ 61
Calibration type
Parameters in
[SENSe<Ch>:]CORRection:
COLLect:METHod:DEFine
Available error terms (depending on port num-
bers)
One-port normalization (reflection)
using an open or a short standard
REFL, RSHort
'REFLTRACK'
Full one port ("Refl OSM")
FOPort
'DIRECTIVITY', 'SRCMATCH', 'REFLTRACK'
Two-port normalization
FRTRans
'TRANSTRACK'
One path two port
OPTPort
'DIRECTIVITY', 'SRCMATCH', 'REFLTRACK',
'TRANSTRACK'
TOSM
TOSM
'DIRECTIVITY', 'SRCMATCH', 'REFLTRACK',
'LOADMATCH', 'TRANSTRACK'
TOM, TSM, TRM, TRL, TNA
TOM | TSM | TRM | TRL | TNA
'DIRECTIVITY', 'SRCMATCH', 'REFLTRACK',
'LOADMATCH', 'TRANSTRACK' (for reading and
writing)
'G11' ... 'G22' and 'H11', 'H12', 'H22' (for reading
only; the 'H21' matrix elements are normalized to
1)
[SENSe<Ch>:]CORRection:CONNection
<ConnectorName>[, <Mode>[, <Gender>[,
<RelPermittivity>[, <RefImpedance>]]]]
Configures the user-defined connector types.
Suffix:
<Ch>
.
Channel number
Parameters:
<ConnectorName>
Name of the user-defined connectors, string parameter.
<Mode>
TEM | WGUide
Transverse electric or waveguide type propagation mode.
<Gender>
GENDer | NGENder
GENDer
–
polar connector type (m/f)
NGENder
–
sexless connector type
<RelPermittivity>
Relative permittivity
Range:
0.0000000001 to 1000.
<RefImpedance>
For TEM type connectors: reference impedance in Ω (without
unit), For waveguide (
WGUide
) type connectors: cutoff frequency
in Hz (without unit).
Range:
Ref. impedance: 1µΩ to 1000 MΩ. Cutoff fre-
quency: 0 Hz to 1000 GHz.
*RST:
-
SCPI command reference
www.allice.de
Allice Messtechnik GmbH