GUI reference
R&S
®
ZND
560
User Manual 1173.9557.02 ─ 61
Note that if the "Operators Check" definition uses custom settings, then the
opch
file
possibly references external files. E.g., for
with custom limit lines, the
generated "OperatorsCheck XML" file
<filename>.opch
references external limit
line files
<filename>_opch<i>.line
. To be able to recall the operator check defini-
tion, keep all these files in one directory and do not rename the external files.
Selftest page
In case you selected "Selftest" on the
, the next step in the "Operators
ZND firmware.
Execute/Repeat
Executes the selftest for the first time in this "Operators Check" wizard session, or
repeats it.
Remote command:
Power Check page
In case you selected "Power Check" on the
port power checks from this page.
During the port power check, the firmware performs a 201-point-linear frequency
sweep over the whole frequency range of the analyzer.
By default, the source power is set to -10
dBm and the firmware checks whether the
measured a- and b-waves are at most 3
dBm above or below this value.
Setup softtool