Command reference
R&S
®
ZND
863
User Manual 1173.9557.02 ─ 61
Example:
HCOP:PAGE:WIND SING; :HCOP
Select one diagram per page and start printing.
7.3.8
INITiate commands
The
INITiate...
commands control the initiation of the trigger system and define
the scope of the triggered measurement.
..............................................................................................863
...............................................................................................864
.................................................................................................864
INITiate<Ch>[:IMMediate][:DUMMy]
.................................................................................865
INITiate<Ch>[:IMMediate]:SCOPe
................................................................................... 865
............................................................................................... 865
INITiate<Ch>:CONTinuous
<Boolean>
Qualifies whether the analyzer measures in single sweep or in continuous sweep
mode.
Suffix:
<Ch>
.
Channel number.
This suffix is ignored in the "ZVR" and "ZVABT" compatibility
modes (
'ZVR' | 'ZVABT'
).
Parameters:
<Boolean>
ON - the analyzer measures continuously, repeating the current
sweep.
OFF - the measurement is stopped after the number of sweeps
defined via
initiates a new measurement cycle.
*RST:
ON
Example:
*RST; :CALC2:PAR:SDEF 'TRC2','S11'
DISPlay:WINDow:TRACe2:FEED 'Trc2'
Reset the analyzer to create the default channel no. 1 and
default trace. Create a second trace and display the trace. Both
traces are measured continuously.
INIT1:CONT OFF
Activate single sweep mode for the first channel. The measure-
ment in channel no. 1 is stopped after the current sweep. Chan-
nel no. 2 is still measured continuously.
INIT1
Activate a new (single) sweep in channel no. 1. Channel no. 2 is
still measured continuously.
INIT:CONT:ALL OFF
Activate single sweep mode for all channels. The measurement
in channel no. 2 is also stopped after the current sweep.
INIT:ALL
Re-start a single sweep in both channels.
SCPI command reference