Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "MUF121" |
"MUF221" | "KFAC21"
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "MUF121" |
"MUF221" | "KFAC21"
Balanced Ports...
Opens a dialog to define a balanced port configuration.
See
Chapter 5.2.1.3, "Balanced Ports dialog"
5.2.9
Power Sensor tab
Allows you to set up and perform measurements using external power sensors.
The standard test setup for a "Power Sensor" measurement involves one analyzer
source port and a power sensor. The power sensor is connected to the VNA (e.g. to
the analyzer's USB port) and provides scalar wave quantity results. See
ter 4.7.19, "External power meters"
Power Meter
Shows a list of all power meters that have been properly configured. See
The bordered label below displays the type and serial number of the selected power
meter.
Remote command:
CALCulate<Ch>:PARameter:MEASure
'TraceName', 'PmtrD1 | ...
CALCulate<Ch>:PARameter:SDEFine
'TraceName', 'PmtrD1 | ...
Auto Zero
Initiates an automatic zeroing procedure of the selected power meter.
The power meter must be disconnected from the RF power; see
on page 219. A message indicates that zeroing is finished.
Remote command:
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:AZERo
Source Port
Selects one of the available test ports of the analyzer as a source of the stimulus sig-
nal.
Note:
For a unidirectional R&S
on page 80) only port 1 may serve as source and hence
this control is disabled.
Remote command:
CALCulate<Ch>:PARameter:MEASure
'TraceName', 'PmtrD1 | ...
CALCulate<Ch>:PARameter:SDEFine
'TraceName', 'PmtrD1 | ...
Meas softtool