Annexes
R&S
®
ZND
1203
User Manual 1173.9557.02 ─ 61
Signal
*RST Configuration
/OUTPUT 1, /OUTPUT 2
"High"
/SWEEP END
"High"
/PASS_FAIL
"High"
SCPI commands:
(resets all configurable signals)
12.3.4.4
Timing of control Signals
The timing of the essential measurement control signals is shown in the figure below.
The duration of the shaded time intervals depends on the measurement settings.
Figure 12-9: Timing of measurement control signals
The figure above corresponds to the default configuration of the /PASS FAIL signal.
For alternative configurations, refer to the documentation of the
CONTrol:HANDler:PASSfail...
on page 819. The figure contains the following pulse durations and response
times.
Time
Description
Value
T1
Pulse duration of /EXTERNAL TRIGGER
Minimum value: 1 μs
T2
Response time of /PASS FAIL STROBE
1 μs
T3
Pulse duration of /PASS FAIL STROBE
1 μs
T4
Pulse duration of /SWEEP END
12 μs
Interfaces and connectors