GUI reference
R&S
®
ZND
580
User Manual 1173.9557.02 ─ 61
simply use "Copy from local network" to copy the "Subnet Mask" and "Standard Gate-
way" from the analyzer's IP settings.
5.17.5.4
Switch Matrix RF Connections dialog
The "Switch Matrix RF Connections" dialog allows you to define the RF connections
between the VNA and the switch matrices and the RF connections to the (DUT) test
ports.
Access:
SYSTEM – [SETUP] > "External Ports" > "RF Connections..."
The following configuration rules apply:
●
Every VNA port must either be connected to a matrix port or directly assigned to a
DUT test port
●
Every matrix that was added to the connection configuration must be connected to
the VNA by at least one RF connection
●
Every VNA port that is not connected to a matrix must be assigned to a test port
●
The test ports must be numbered consecutively, starting with 1
Configuration errors are displayed in balloon popups appearing at the lower left corner
of the dialog.
Background information
Chapter 4.7.20, "External switch matrices"
The definition of the RF connections typically involves the following steps:
1. "Add" one or more switching matrices
These matrices must have been registered before; see
on page 575 for details.
Setup softtool
www.allice.de
Allice Messtechnik GmbH