GUI reference
R&S
®
ZND
449
User Manual 1173.9557.02 ─ 61
If "Number of Alternated Blocks" is set to 1, "Block Alternated" is equivalent to "Alterna-
ted" mode. If it is set to the
, "Block Alternated" is equivalent to
"Chopped" mode. By selecting a suitable
, you can minimize the total
sweep duration.
Note:
●
If the number of sweep points is not divisible by the number of blocks, the first
block is extended by the excessive number sweep points.
●
For segmented sweeps the "Block Alternated"/"Number of Alternated Blocks" logic
is applied to the individual sweep segments.
Remote command:
[SENSe<Ch>:]COUPle BLOCk, <Number of Alternated Blocks>
5.12.4
Pwr Cal Settings tab
Replicates the "Pwr Cal Settings" tab of the "Cal" softtool (see
5.13
Offset Embed softtool
The "Offset Embed" softtool allows you to define a length offset and loss for each test
port. The offset compensates for the known length and loss of (non-dispersive and per-
fectly matched) transmission lines between the calibrated reference plane and the
DUT. It also contains advanced functions for deembedding/embedding the DUT from/
into more general physical/virtual (matching) networks placed between the calibrated
reference plane and the DUT.
Access:
CHANNEL – [OFFSET
EMBED]
Background information
Refer to
Chapter 4.6, "Offset parameters and de-/embedding"
5.13.1
Offset Embed dock widget
On activating a tab in the "Offset Embed" softtool, a dock widget is displayed beneath
the trace area, whose content pane is synchronized with the selected softtool tab.
Access:
CHANNEL – [OFFSET
EMBED]
5.13.1.1
Overview panel
Shows an overview of the overall calculation flow and provides quick access to the
"Offset Embed" functions.
Access:
CHANNEL – [OFFSET
EMBED] > "Overview"
Offset Embed softtool
www.allice.de
Allice Messtechnik GmbH